Which Medical Device › Closure devices
Angio-Seal vs Mynxgrip Vascular Closure Device
Angio-Seal by Terumo and Mynxgrip Vascular Closure Device by Cordis, side by side.
Which Medical Device summary
Angio-Seal (Terumo) and Mynxgrip (Cordis) are both vascular closure devices used to achieve haemostasis at femoral access sites after catheter-based procedures. The main difference lies in their closure mechanism and location: Angio-Seal uses an intravascular bioabsorbable anchor paired with an extravascular collagen sponge, sandwiching the arteriotomy via a suture, while Mynxgrip relies solely on an extravascular PEG hydrogel sealant delivered through the existing sheath, with no intravascular component or suturing.
- Angio-Seal may suit operators wanting an anchor-and-sponge mechanism with a long track record, compatible with 6F or 8F sheaths for arterial closure.
- Mynxgrip may suit cases needing closure of both femoral arterial and venous access sites, using smaller 5–7F sheaths, with a purely extravascular, suture-free approach.
On WMD, Angio-Seal has 5 listed papers, including a meta-analysis and randomised trials comparing it with StarClose and manual compression, alongside 4.5/5 ratings from 4 clinicians. Mynxgrip has 7 listed papers, including several randomised trials (e.g. CLOSE-UP III, PANDA) and case reports covering niche applications, with 4.5/5 ratings from 2 clinicians.
Summary checked by the WMD editor, 2026-11-29.
Side by side
| Angio-Seal | Mynxgrip Vascular Closure Device | |
|---|---|---|
| Manufacturer | Terumo | Cordis |
| Clinician rating | 4.5 / 5 (4 ratings) | 4.5 / 5 (2 ratings) |
| Published papers on WMD | 5 | 7 |
| Price benchmark (AU) | A$290 | — |
| Closure Location | Intravascular + extravascular (bioabsorbable anchor seats against the inside of the arteriotomy; collagen sponge sits extravascularly, sandwiching the arteriotomy between the two) | Extravascular |
| Closure Mechanism | Collagen plug + anchor — bioabsorbable polymer anchor and collagen sponge tethered by an absorbable self-tightening suture (STS) | Extravascular sealant — "Grip" polyethylene glycol (PEG) hydrogel that actively adheres to the artery/vein wall and expands to fill the tissue tract; delivered via balloon catheter, no sutures, clips, or metal implant |
| Compatible Sheath Size | 6F or 8F (device size must match: 6F device for ≤6F sheath, 8F device for ≤8F sheath) | 5F, 6F, or 7F |
| Intravascular Component | Yes — bioabsorbable polymer anchor, remains until absorbed | — |
| Implant Material | Bioabsorbable polymer anchor, collagen sponge, and absorbable suture | Polyethylene glycol (PEG) hydrogel sealant |
| Bioabsorbable vs Permanent | Bioabsorbable — anchor begins hydrating/softening 24-36 hours after deployment and is ~95% absorbed by 42 days; all components absorbed within 60-90 days | Bioabsorbable — sealant dissolves within 30 days; manufacturer states nothing is permanently left behind |
| Access Vessel / Site | Common femoral artery | Femoral artery and femoral vein |
| Number of Devices Required | — | 1 — single balloon catheter with integrated sealant, plus a 10 mL locking syringe |
| Pre-Close or Post-Close | — | Post-close — deployed through the existing procedural sheath after the interventional procedure |
| Anatomical Restrictions | — | Contraindicated if the puncture site is above the most inferior border of the inferior epigastric artery (arterial use) and/or above the inguinal ligament by osseous landmarks — risk of retroperitoneal hematoma/bleed |
Prices are published national reimbursement or contract benchmarks, not what a hospital pays.
About each device
Angio-Seal
Terumo · Closure devices
Angio-seal is a medical device indicated for use in closing and reducing time to hemostasis at the femoral arterial puncture site in patients who have undergone diagnostic angiography procedures or interventional procedures (minimal artery diameter of 4mm). The 6F Angio-Seal device is for closure after usage of a 6 French or smaller procedural sheath, while the 8F Angio-Seal device is for closure after usage of an 8 French or smaller procedural sheath.
Mynxgrip Vascular Closure Device
Cordis · Closure devices
The MYNXGRIP® Vascular Closure Device
Published evidence
Angio-Seal
- Arterial closure devices versus manual compression for femoral haemostasis in interventional radiological procedures: a systematic review and meta-analysis. Cardiovasc Intervent Radiol 2011
- Closure of the femoral artery after cardiac catheterization: a comparison of Angio-Seal, StarClose, and manual compression.. Catheter Cardiovasc Interv 2008
- A comparative evaluation of arterial blood flow and the healing response after femoral artery closure using angio-seal STS Plus and StarClose in a porcine model.. J Interv Cardiol 2008
- Correspondence to article: Closure of the femoral artery after cardiac catheterization--a comparison of Angio-Seal, StarClose, and manual compression.. Catheter Cardiovasc Interv 2008
- A randomized controlled trial comparing StarClose and AngioSeal vascular closure devices in a district general hospital.. Int J Clin Pract 2008
Mynxgrip Vascular Closure Device
- Comparison of MynxGrip vascular closure device and manual compression for closure after femoral access angiography: a randomized controlled trial: the closure devices used in every day practice study, CLOSE-UP III trial. BMC cardiovascular disorders 2022
- Use of the MynxGrip vascular closure device in patients undergoing interventional diagnosis or treatment: The PANDA multicenter, open-label, randomized controlled trial. The journal of vascular access 2024
- MynxGrip® vascular closure device versus manual compression for hemostasis of percutaneous transfemoral venous access closure: Results from a prospective multicenter randomized study. Cardiovascular revascularization medicine : including molecular interventions 2018
- MynxGrip vascular closure device use in pediatric neurointerventional procedures. Journal of neurosurgery. Pediatrics 2018
- Feasibility of Mynxgrip®-Assisted Percutaneous Transhepatic Portal Venous Access Closure. Cardiovascular and interventional radiology 2020
- Endovascular foreign body removal of a MynxGrip polyethylene glycol (PEG) sealant embolus using a cerebral stent retriever. CVIR endovascular 2021
Full details: Angio-Seal · Mynxgrip Vascular Closure Device